| | World wide/Two-stage/withoutthermal protector / with cord,plug,andswitch | 
 
 |  | |  | Itisatype applicable for the voltages all the world, foramulti-voltagemotor is equipped to GLD series. 2 types,forsingle/three phase,are available. 
 As it is applicable for various voltages by a simplewiringchange,there is no necessity to change the motor for useinaccordance witha voltage in the countries where the pumpisused.
 
 For the single phase, it is used at 100V-120V or200V-240V.Thefunctional performance such as gas-ballast mechanism,oilback-flowprevention mechanism, large-sized oil-level gauge, etc.isequal toGLD Series. It isCEself-declaredproducts.
 | 
 | 
 
 | |  Specifications | 
 | |  | Unit | 50Hz | 60Hz |  | ActualPumpingSpeed | L/min. | 135 | 162 |  | Ultimate pressure | Pa | 6.7 x 10-2(6.7×10-1)(G.V.closed)( 6.7 )(G.V.open)
 |  | Motor |  | Three phase, 400W, 4 poles, Multiple-range motor AC200-240V/380-460V
 |  | Full load current | A | 2.1(200V), 2.5 (240V), 1.3 (380V), 1.4 (415V)
 | 2.0(200V), 2.0 (240V), 1.1 (380V), 1.2 (460V)
 |  | Oil capacity | mL | 1,000 |  | Recommended oil |  | SMR-100 |  | Weight | kg | 23 |  | Inlet, outlet pipe diameter | mm | KF-25 |  | Ambient operation temp.range | °C | 7 – 40 |  | Dimensions | mm | 170(W) x 485.5(L)xabout240(H) | 
 | 
 | 
 | *Ultimate pressure is measured withMacleodgauge.In ( ) shows the measurements withPiranigauge. | 
 
 |  Caution | 
 | GLDSeries,compared with G Series, is markedly upgraded. For thesakeofhaving the maximum performance, in addition to notes forGseries,be careful of also the followings. |  | 1. | GLDseriesiscapable of reaching to a high vacuum area (10-2 Pa). Flanges, vessels usable at a high vacuumarerequired. |  | 2. | Anoil-sealed rotary pump is not recommended forcontinuousoperationnear atmospheric pressure. | 
 
 |  Outlinedrawing | 
    
 
 |  PumpingSpeedCurves | 
  
 
 |  Option | 
 Optionalsuctionport
 | |   | Size | Note |  | KF-25 suctionport | KF-25X(50X50) | Attachedtostandardtype |  | Suctionpipeforhose | Dia.8 | Option. 50X50adapter isneeded,usethe suction pipe /50X50 adapter |  | Suctionpipeforhose | Dia.12 | Option. 50X50adapter isneeded,usethe suction pipe /50X50 adapter |  | Suctionpipeforhose | Dia.15 | Option. 50X50adapter isneeded,usethe suction pipe /50X50 adapter |  | Suctionpipeforhose | Dia.18 | Option. 50X50adapter isneeded,usethe suction pipe /50X50 adapter |  | Suctionpipeforhose | Dia.22 | Option. 50X50adapter isneeded,usethe suction pipe /50X50 adapter |  | Suctionpipeforhose | Dia.27 | Option |  | Suctionpipeforhose | Dia.30 | Option |  | Suctionpipeforhose | Dia.34 | Option |  | A typesuctionpipe | PT1/4-L20,male | Option. 50X50adapter isneeded,usethe suction pipe /50X50 adapter |  | B typesuctionpipe | PT1/4-L20,female | Option. 50X50adapter isneeded,usethe suction pipe /50X50 adapter |  | C typesuctionpipe | PT1/4-L10,male | Option. 50X50adapter isneeded,usethe suction pipe /50X50adapter | 
 | 
 Optional exhaust port
 
 | |   | Size | Note |  | Exhaust port cap | - | Attachedtostandardtype |  | Exhaustpipeforhose | Dia.27 | Option |  | KF-25 exhaustport | KF-25 | Option | 
 | 
 Trap for exhaust side
 
 | |   | Model | Note |  | Oil-mist trap | OMT-200A | Option |  | In-linetypeoil-misttrap | OMI-200 | Option | 
 | 
 Trap for suction side
 | |   | Model | Note |  | For-line trap | OFI-200C | Prevent counter flowofoildiffusion into vacuum chamber, Need KF-25clampforattaching |  | For-line trap | OFI-200V | Prevent mixing powder orliquidintothe pump, Need KF-25clumpforattaching | 
 | 
 |