Model | CHW700 | CHS700 |
|---|
Performance
| Operating Water | For city water circulation | For pure water circulation | Circulation Type | Closed systemcirculationbypump | Operating Temperature Range | 10°C to Room temp. | Tem. Adjustment Accuracy | ±1.5 to 2.5°C | Cooling Capacity | Approx. 1,000W (860kcal)atLiquidtemp. 20°C, at Room temp. 20°C | Components
| Temp. Control System | Refrigerator On/Off control | Temp. Sensor | T-thermocouple | Temp. Setting, Display | Digital | Refrigerator •Coolant | Air cooling 675W, R407C | Circulation Pump | Magnet pump 65W x 2pcs. | Pump Max. Flow Rate | 22/25L/min. (50/60Hz) | Pump Max. Lift | 10/13.5m (50/60Hz) | SurroundingOperationTemp.Range | 5 to 30°C | Circulation Nozzle | Rc3/8 Female screw | SafetyDevice
| Safety Countermeasures | Self diagnosisfunctions,Electricleakage breaker with over current protection,Refrigeratoroverload relay, Refrigerator pressuredetector,Refrigeratorprotection’s delay timer function, Floatswitch,Circulation pumpprotection by-pass | Other Function | Drain cock, Dustpreventionfilter,Refrigerator pressure indicator, Automaticstopoperation,Automatic start operation, Key lock function,Calibrationoff setfunction, Breakout protection function, Temp.outputterminal(4-20mA). | Accessories | Drain hose 0.5m x 1pc. | Standard
| Water Bath Dimension | ID300xH236mm | Water Bath Capacity | 14L | Cooling Coil | Copper-nickel plated | SUS304 | External Dimension | W380 x D460 x H1,050 mm | Power Source | AC100V 50/60Hz 12A | Weight | Approx.55kg |
|