Model
|
CHW700
|
CHS700
|
Performance
|
Operating Water
|
For city water circulation
|
For pure water circulation
|
Circulation Type
|
Closed systemcirculationbypump
|
Operating Temperature Range
|
10°C to Room temp.
|
Tem. Adjustment Accuracy
|
±1.5 to 2.5°C
|
Cooling Capacity
|
Approx. 1,000W (860kcal)atLiquidtemp. 20°C, at Room temp. 20°C
|
Components
|
Temp. Control System
|
Refrigerator On/Off control
|
Temp. Sensor
|
T-thermocouple
|
Temp. Setting, Display
|
Digital
|
Refrigerator •Coolant
|
Air cooling 675W, R407C
|
Circulation Pump
|
Magnet pump 65W x 2pcs.
|
Pump Max. Flow Rate
|
22/25L/min. (50/60Hz)
|
Pump Max. Lift
|
10/13.5m (50/60Hz)
|
SurroundingOperationTemp.Range
|
5 to 30°C
|
Circulation Nozzle
|
Rc3/8 Female screw
|
SafetyDevice
|
Safety Countermeasures
|
Self diagnosisfunctions,Electricleakage breaker with over current protection,Refrigeratoroverload relay, Refrigerator pressuredetector,Refrigeratorprotection’s delay timer function, Floatswitch,Circulation pumpprotection by-pass
|
Other Function
|
Drain cock, Dustpreventionfilter,Refrigerator pressure indicator, Automaticstopoperation,Automatic start operation, Key lock function,Calibrationoff setfunction, Breakout protection function, Temp.outputterminal(4-20mA).
|
Accessories
|
Drain hose 0.5m x 1pc.
|
Standard
|
Water Bath Dimension
|
ID300xH236mm
|
Water Bath Capacity
|
14L
|
Cooling Coil
|
Copper-nickel plated
|
SUS304
|
External Dimension
|
W380 x D460 x H1,050 mm
|
Power Source
|
AC100V 50/60Hz 12A
|
Weight
|
Approx.55kg
|
|