| | ADP200 | ADP300 |
|---|
Performance
|
| System | Vacuumdryingbydecompressed chamber direct heating |
| Operatingtemperaturerange | Roomtemp.+5to240°C |
| Operatingpressurerange | 101 to0.1kPa(760to 1 Torr) at absolute pressure |
| Temp.controlaccuracy | ±1.5°C(at140°C) |
| Temp.distributionaccuracy | ±10°C(at240°C) |
| Max.temp.reachingtime | Approx. 70 min. | Approx. 100 min. |
Components |
| InteriorMaterial | StainlesssteelSUS304 |
| ExteriorMaterial | Coldrolledsteelplate with baked-on melamine resin finish |
| Door | Singleswingdoor |
| Heatinsulatingmaterial | Rock wool |
| Heatingmethod | Decompressedchamber direct heating |
| Heater | Mica heater, 0.68kW | Mica heater, 1.05kW |
| Vacuumgauge | Bourdontubetype,0to -0.1 MPa (Gauge pressure) |
| Observationwindow | Temperedglass(12mmthickness) and polycarbonate resin plate |
Controller |
| Temp.controlmethod | PlDcontrolbymicroprocessor |
| Temp.settingmethod | Digitalsettingwith key |
| Tempdisplaymethod | Digitaldisplay |
| Timer | 1 min. to99Hrs.59min. and 100 Hrs. to 999.5 Hrs. Digital display |
| Min.division | 1min.or1Hrs. |
| Operationfunction | Fixedtemperatureoperation, Auto-startoperation,Auto-stopoperation |
| Additionalfunctions | Calendartimer(max.24 Hrs.), Integration time (max. 49999Hrs.),Timedisplay |
| Heatercircuitcontrol | Triaczero-crosscontrol |
| Temp.sensor | K-thermocouple(double sensor) |
SafeDevice |
| Safetycountermeasures | Selfdiagnosisfunctions(Sensor, Heater, Triac,automaticoverheatingprevention),Independent overheating prevention,Key lockfunction,Electricleakage breaker |
Standard |
Internaldimensions(W×D×Hmm) External dimensions(W×D×Hmm) Internal capacity | 200×250×200 400×411×565 10 liters | 300×300×300 500×463×669 27 liters |
| Shelfreststepnumber | 2 steps | 3 steps |
| Shelfrestpitch | 63mm | 71mm |
| Vacuumport | 18mm O.D. |
| Powersource50/60Hz | 100V/220V, single phase, 8A/3.7A | 100V/220V, single phase, 12A/5.5A |
| Weight | Approx. 30 kg | Approx. 55 kg |
Accessories |
| Shelfplate | Punchedstainlesssteel, 2 pcs |
OptionalAccessories |
| Additional shelf plate/Temp.outputterminal / Temp. program function / Adaptor forvacuumhose/ Vacuumhose etc. |